digital-brain/content/zettels/interferometers.md

77 lines
4.0 KiB
Markdown
Raw Blame History

This file contains ambiguous Unicode characters

This file contains Unicode characters that might be confused with other characters. If you think that this is intentional, you can safely ignore this warning. Use the Escape button to reveal them.

+++
title = "Interferometers"
author = ["Thomas Dehaeze"]
draft = false
+++
Backlinks:
- [Position Sensors]({{< relref "position_sensors" >}})
Tags
: [Position Sensors]({{< relref "position_sensors" >}})
## Manufacturers {#manufacturers}
| Manufacturers | Links | Country |
|---------------|----------------------------------------------------------------------------------------------------------|-------------|
| Attocube | [link](http://www.attocube.com/) | Germany |
| Zygo | [link](https://www.zygo.com/?/met/markets/stageposition/zmi/) | USA |
| Smaract | [link](https://www.smaract.com/interferometry) | Germany |
| Qutools | [link](https://www.qutools.com/qudis/) | Germany |
| Renishaw | [link](https://www.renishaw.com/en/fibre-optic-laser-encoder-products--6594) | UK |
| Sios | [link](https://sios-de.com/products/length-measurement/laser-interferometer/) | Germany |
| Keysight | [link](https://www.keysight.com/en/pc-1000000393%3Aepsg%3Apgr/laser-heads?nid=-536900395.0&cc=FR&lc=fre) | USA |
| Optics11 | [link](https://optics11.com/) | Netherlands |
## Environmental Units {#environmental-units}
<div class="table-caption">
<span class="table-number">Table 1</span>:
Characteristics of Environmental Units
</div>
| | Temperature (\\(\pm\ ^oC\\)) | Pressure (\\(\pm\ hPa\\)) | Humidity \\(\pm\\% RH\\) | Wavelength Accuracy (\\(\pm\ \text{ppm}\\)) |
|-----------|------------------------------|---------------------------|--------------------------|---------------------------------------------|
| Attocube | 0.1 | 1 | 2 | 0.5 |
| Renishaw | 0.2 | 1 | 6 | 1 |
| Picoscale | 0.2 | 2 | 2 | 1 |
## Interferometer Precision {#interferometer-precision}
([Jang and Kim 2017](#orgc0eaaa4))
<a id="orge2a3743"></a>
{{< figure src="/ox-hugo/position_sensor_interferometer_precision.png" caption="Figure 1: Expected precision of interferometer as a function of measured distance" >}}
## Sources of uncertainty {#sources-of-uncertainty}
Sources of error in laser interferometry are well described in ([Ducourtieux 2018](#org2c23555)).
It includes:
- Laser Source Stability
- Variation of refractive index of air, which is dependent of:
- Temperature: \\(K\_T \approx 1 ppmK^{-1}\\)
- Pressure: \\(K\_P \approx 0.27 ppm hPa^{-1}\\)
- Humidity: \\(K\_{HR} \approx 0.01 ppm \% RH^{-1}\\)
- These errors can partially be compensated using an environmental unit.
- Air turbulence (Figure [2](#org92b4926))
- Non linearity
<a id="org92b4926"></a>
{{< figure src="/ox-hugo/interferometers_air_turbulence.png" caption="Figure 2: Effect of air turbulences on measurement stability" >}}
## Bibliography {#bibliography}
<a id="org2c23555"></a>Ducourtieux, Sebastien. 2018. “Toward High Precision Position Control Using Laser Interferometry: Main Sources of Error.” <https://doi.org/10.13140/rg.2.2.21044.35205>.
<a id="orgc0eaaa4"></a>Jang, Yoon-Soo, and Seung-Woo Kim. 2017. “Compensation of the Refractive Index of Air in Laser Interferometer for Distance Measurement: A Review.” _International Journal of Precision Engineering and Manufacturing_ 18 (12):188190. <https://doi.org/10.1007/s12541-017-0217-y>.