7.8 KiB
+++ title = "Interferometers" author = ["Dehaeze Thomas"] draft = false category = "equipment" +++
- Tags
- [Position Sensors]({{< relref "position_sensors.md" >}}), [Optics]({{< relref "optics.md" >}})
Manufacturers
Manufacturers | Country |
---|---|
Attocube | Germany |
Zygo | USA |
Smaract | Germany |
Qutools | Germany |
Renishaw | UK |
Sios | Germany |
Keysight | USA |
Optics11 | Netherlands |
Prodrive | Netherlands |
Agito |
Reviews
(Ducourtieux 2018, 2018; Bobroff 1993, 1993; Thurner et al. 2015, 2015; Loughridge and Abramovitch 2013)
Effect of Refractive Index - Environmental Units
The measured distance is proportional to the refractive index of the air that depends on several quantities as shown in Table 1 (Taken from (Thurner et al. 2015)).
Physical Value | Refractive Index Sensitivity | Value |
---|---|---|
Temperature \(T\) | \(dn/dT\ (K^{-1})\) | \(-9.32\cdot 10^{-7}\) |
Pressure \(p\) | \(dn/dp\ (mbar^{-1})\) | \(2.70\cdot 10^{-7}\) |
Humidity \(h\) | \(dn/dh\ (\text{%}^{-1})\) | \(-8.72\cdot 10^{-9}\) |
\(\text{CO}_2\) content \(x_c\) | \(dn/dx_c\ (ppm^{-1})\) | \(1.42\cdot 10^{-10}\) |
Wavelength \(\lambda\) | \(dn/d\lambda\ (nm^{-1})\) | \(-8.59\cdot 10^{-10}\) |
In order to limit the measurement uncertainty due to variation of air parameters, an Environmental Unit can be used that typically measures the temperature, pressure and humidity and compensation for the variation of refractive index in real time.
Typical characteristics of commercial environmental units are shown in Table 2.
Temperature (\(\pm\ ^oC\)) | Pressure (\(\pm\ hPa\)) | Humidity \(\pm\% RH\) | Wavelength Accuracy (\(\pm\ \text{ppm}\)) | |
---|---|---|---|---|
Attocube | 0.1 | 1 | 2 | 0.5 |
Renishaw | 0.2 | 1 | 6 | 1 |
Picoscale | 0.2 | 2 | 2 | 1 |
Interferometer Precision
Figure 1 shows the expected precision as a function of the measured distance due to change of refractive index of the air (taken from (Jang and Kim 2017)).
{{< figure src="/ox-hugo/position_sensor_interferometer_precision.png" caption="<span class="figure-number">Figure 1: Expected precision of interferometer as a function of measured distance" >}}
Sources of uncertainty
Sources of error in laser interferometry are well described in (Ducourtieux 2018).
It includes:
- Laser Source Stability
- Variation of refractive index of air, which is dependent of:
- Temperature: \(K_T \approx 1 ppmK^{-1}\)
- Pressure: \(K_P \approx 0.27 ppm hPa^{-1}\)
- Humidity: \(K_{HR} \approx 0.01 ppm \% RH^{-1}\)
- These errors can partially be compensated using an environmental unit.
- Air turbulence (Figure 2)
- Non linearity
{{< figure src="/ox-hugo/interferometers_air_turbulence.png" caption="<span class="figure-number">Figure 2: Effect of air turbulences on measurement stability" >}}