title = "An instrument for 3d x-ray nano-imaging"
|
||
|
author = ["Thomas Dehaeze"]
|
||
|
draft = false
|
||
|
+++
|
||
|
|
||
|
Tags
|
||
|
: [Nano Active Stabilization System]({{< relref "nano_active_stabilization_system" >}}), [Positioning Stations]({{< relref "positioning_stations" >}})
|
||
|
|
||
|
Reference
|
||
|
: <sup id="66ab0e7602a1dedda963d7da60533b0d"><a href="#holler12_instr_x_ray_nano_imagin" title="Holler, Raabe, Diaz, Guizar-Sicairos, , Quitmann, Menzel \& Bunk, An Instrument for 3d X-Ray Nano-Imaging, {Review of Scientific Instruments}, v(7), 073703 (2012).">(Holler {\it et al.}, 2012)</a></sup>
|
||
|
|
||
|
Author(s)
|
||
|
: Holler, M., Raabe, J., Diaz, A., Guizar-Sicairos, M., Quitmann, C., Menzel, A., & Bunk, O.
|
||
|
|
||
|
Year
|
||
|
: 2012
|
||
|
|
||
|
Instrument similar to the NASS.
|
||
|
Obtain position stability of 10nm (standard deviation).
|
||
|
|
||
|
<a id="orgba4a339"></a>
|
||
|
|
||
|
{{< figure src="/ox-hugo/holler12_station.png" caption="Figure 1: Schematic of the tomography setup" >}}
|
||
|
|
||
|
- **Limited resolution due to instrumentation**:
|
||
|
The resolution of ptychographic tomography remains above 100nm due to instabilities and drifts of the scanning systems.
|
||
|
- **Need of a Metrology System**:
|
||
|
|
||
|
> To achieve positioning accuracy and stability in the nanometer range, one cannot rely on the position encoders built into individual positioning stages.
|
||
|
> A precise exteroceptive measurement of the relative position of the optical elements with respect to the sample is mandatory.
|
||
|
> Thus, thermal drifts and parasitic motions can be measured and compensated for.
|
||
|
- **Interferometer System Concept**:
|
||
|
The sample is aligned with the X-ray with the XYZ piezo stage.
|
||
|
As a result, the metrology sphere will be usually off center with respect to the rotation axis of the spindle.
|
||
|
That implies that the laser will not propagate back to the interferometer at all rotation angles.
|
||
|
A position sensitive detector (PSD) is used, it provides a measurement of the position of the sphere in the plane perpendicular to the laser.
|
||
|
The interferometer is positionned on top of a translation stage. The PSD information is used to close the loop so that the interferometer follows the displacement of the metrology sphere.
|
||
|
- **Feedback Loop**: Using the signals from the 2 interferometers, the loop is closed to compensate low frequency vibrations and thermal drifts.
|
||
|
|
||
|
# Bibliography
|
||
|
<a id="holler12_instr_x_ray_nano_imagin"></a>Holler, M., Raabe, J., Diaz, A., Guizar-Sicairos, M., Quitmann, C., Menzel, A., & Bunk, O., *An instrument for 3d x-ray nano-imaging*, Review of Scientific Instruments, *83(7)*, 073703 (2012). http://dx.doi.org/10.1063/1.4737624 [↩](#66ab0e7602a1dedda963d7da60533b0d)
|