#+TITLE:Measurement of the ID31 Micro-Station

This web-page gathers all the measurements done on the ID31 Micro Station.

Measurements of the dynamics of the station

Measurement 1

Link to the analysis

Notes

Date 2017-11-17
Sensors Accelerometers
Excitation Instrumented Hammer
Location Laboratory

Dynamics of the station is evaluated using instrumented hammer and accelerometers fixed on each stage.

Goal

  • Obtain a first estimation of resonance frequencies

Results

  • Resonances have been identified at 45Hz and 75Hz
  • However, the quality of the measurements are bad at low frequency
  • New measurements should be done with Geophones

Measurement 2

Link to the analysis

Notes

Date 2018-01-12
Sensors Geophones
Excitation Instrumented Hammer
Location Laboratory
Notes Unglued Granite

Goal

Obtain better coherence at low frequency.

Results

  • Resonances at 42Hz, 70Hz and 125Hz have been identified
  • The coherence is much better than when using accelerometers

TODO Measurement 3

Link to the analysis

Notes

Date 2018-10-12
Sensors Geophones
Excitation Instrumented Hammer
Location Experimental Hutch

Goal

The station is now installed on the experimental hutch with a glued granite (final location). The station is identified again.

Results