UP | HOME

Voltage Amplifier PD200 - Test Bench

Table of Contents

1 Introduction

The goal of this test bench is to characterize the Voltage amplifier PD200 from PiezoDrive.

The documentation of the PD200 is accessible here.

amplifier_PD200.png

Figure 1: Picture of the PD200 Voltage Amplifier

2 Voltage Amplifier Requirements

Table 1: Requirements for the Voltage Amplifier
  Specification
Continuous Current > 50 [mA]
Output Voltage Noise (1-200Hz) < 2 [mV rms]
Voltage Input Range +/- 10 [V]
Voltage Output Range -20 [V] to 150 [V]
Small signal bandwidth (-3dB) > 5 [kHz]

3 PD200 Expected characteristics

Table 2: Characteristics of the PD200
Characteristics Manual Specification
Input Voltage Range +/- 10 [V] +/- 10 [V]
Output Voltage Range -50/150 [V] -20/150 [V]
Gain 20 [V/V]  
Maximum RMS current 0.9 [A] > 50 [mA]
Maximum Pulse current 10 [A]  
Slew Rate 150 [V/us]  
Noise (10uF load) 0.7 [mV RMS] < 2 [mV rms]
Small Signal Bandwidth (10uF load) 7.4 [kHz] > 5 [kHz]
Large Signal Bandwidth (150V, 10uF) 300 [Hz]  

For a load capacitance of \(10\,\mu F\), the expected \(-3\,dB\) bandwidth is \(6.4\,kHz\) (Figure 2) and the low frequency noise is \(650\,\mu V\,\text{rms}\) (Figure 3).

pd200_expected_small_signal_bandwidth.png

Figure 2: Expected small signal bandwidth

pd200_expected_noise.png

Figure 3: Expected Low frequency noise from 0.03Hz to 20Hz

4 Voltage Amplifier Model

The Amplifier is characterized by its dynamics \(G_a(s)\) from voltage inputs \(V_{in}\) to voltage output \(V_{out}\). Ideally, the gain from \(V_{in}\) to \(V_{out}\) is constant over a wide frequency band with very small phase drop.

It is also characterized by its output noise \(n\). This noise is described by its Power Spectral Density.

The objective is therefore to determine the transfer function \(G_a(s)\) from the input voltage to the output voltage as well as the Power Spectral Density \(S_n(\omega)\) of the amplifier output noise.

As both \(G_a\) and \(S_n\) depends on the load capacitance, they should be measured when loading the amplifier with a \(10\,\mu F\) capacitor.

pd200-model-schematic.png

Figure 4: Model of the voltage amplifier

5 Noise measurement

5.1 Setup

Here are the documentation of the equipment used for this test bench:

The output noise of the voltage amplifier PD200 is foreseen to be around 1mV rms in a bandwidth from DC to 1MHz. If we suppose a white noise, this correspond to an amplitude spectral density:

\begin{equation} \phi_{n} \approx \frac{1\,mV}{\sqrt{1\,MHz}} = 1 \frac{\mu V}{\sqrt{Hz}} \end{equation}

The RMS noise begin very small compare to the ADC resolution, we must amplify the noise before digitizing the signal. The added noise of the instrumentation amplifier should be much smaller than the noise of the PD200. We use the amplifier EG&G 5113 that has a noise of \(\approx 4 nV/\sqrt{Hz}\) referred to its input which is much smaller than the noise induced by the PD200.

The gain of the low-noise amplifier can be increased until the full range of the ADC is used. This gain should be around 1000.

setup-noise-measurement.png

Figure 5: Schematic of the test bench to measure the Power Spectral Density of the Voltage amplifier noise \(n\)

A low pass filter at 10kHz can be included in the EG&G amplifier in order to limit aliasing. An high pass filter at low frequency can be added if there is a problem of large offset.

5.2 Model of the setup

As shown in Figure 6, there are 4 equipment involved in the measurement:

  • a Digital to Analog Convert (DAC)
  • the Voltage amplifier to be measured with a gain of 20 (PD200)
  • a low noise voltage amplifier with a variable gain and integrated low pass filters and high pass filters
  • an Analog to Digital Converter (ADC)

Each of these equipment has some noise:

  • \(q_{da}\): quantization noise of the DAC
  • \(n_{da}\): output noise of the DAC
  • \(n_p\): output noise of the PD200 (what we wish to characterize)
  • \(n_a\): input noise of the pre amplifier
  • \(q_{ad}\): quantization noise of the ADC

noise_meas_procedure.png

Figure 6: Sources of noise in the experimental setup

5.3 Quantization Noise

The quantization noise is something that can be predicted. The Amplitude Spectral Density of the quantization noise of an ADC/DAC is equal to:

\begin{equation} \Gamma_q(\omega) = \frac{q}{\sqrt{12 f_s}} \end{equation}

with:

  • \(q = \frac{\Delta V}{2^n}\) the quantization in [V], which is the corresponding value in [V] of the least significant bit
  • \(\Delta V\) is the full range of the ADC in [V]
  • \(n\) is the number of bits
  • \(f_s\) is the sample frequency in [Hz]
adc = struct();
adc.Delta_V = 20; % [V]
adc.n = 16; % number of bits
adc.Fs = 20e3; % [Hz]
adc.Gamma_q = adc.Delta_V/2^adc.n/sqrt(12*adc.Fs); % [V/sqrt(Hz)]

The obtained Amplitude Spectral Density is 6.2294e-07 \(V/\sqrt{Hz}\).

5.4 Pre Amplifier noise measurement

First, we wish to measure the noise of the pre-amplifier. To do so, the input of the pre-amplifier is shunted such that there is 0V at its inputs. Then, the gain of the amplifier is increase until the measured signal on the ADC is much larger than the quantization noise.

The Amplitude Spectral Density \(\Gamma_n(\omega)\) of the measured signal \(n\) is computed. Finally, the Amplitude Spectral Density of \(n_a\) can be computed taking into account the gain of the pre-amplifier:

\begin{equation} \Gamma_{n_a}(\omega) \approx \frac{\Gamma_n(\omega)}{|G_a(\omega)|} \end{equation}

This is true if the quantization noise \(\Gamma_{q_{ad}}\) is negligible.

noise_measure_setup_preamp.png

Figure 7: Sources of noise in the experimental setup

The gain of the low noise amplifier is set to 50000.

% Hanning window
win = hanning(ceil(0.5/Ts));

% Power Spectral Density
[pxx, f] = pwelch(preamp.Vn, win, [], [], Fs);

% Save the results inside the struct
preamp.pxx = pxx;
preamp.f = f;

The obtained Amplitude Spectral Density of the Low Noise Voltage Amplifier is shown in Figure 8. The obtained noise amplitude is very closed to the one specified in the documentation of \(4nV/\sqrt{Hz}\) at 1kHZ.

asd_preamp.png

Figure 8: Obtained Amplitude Spectral Density of the Low Noise Voltage Amplifier

5.5 PD200 noise measurement

The input of the PD200 amplifier is shunted such that there is 0V between its inputs. Then the gain of the pre-amplifier is increased in order to measure a signal much larger than the quantization noise of the ADC. We compute the Amplitude Spectral Density of the measured signal \(\Gamma_n(\omega)\). The Amplitude Spectral Density of \(n_p\) can be computed taking into account the gain of the pre-amplifier:

\begin{equation} \Gamma_{n_p}(\omega) = \frac{\Gamma_n(\omega)}{|G_a(\omega)|} \end{equation}

And we verify that this is indeed the noise of the PD200 and not the noise of the pre-amplifier by checking that:

\begin{equation} \Gamma_{n_p} \ll \Gamma_{n_a} \end{equation}

noise_measure_setup_pd200.png

Figure 9: Sources of noise in the experimental setup

The measured low frequency noise \(n_p\) of one of the amplifiers is shown in Figure 10. It is very similar to the one specified in the datasheet in Figure 3.

pd200_noise_time_lpf.png

Figure 10: Measured low frequency noise of the PD200 from 0.01Hz to 20Hz

The obtained RMS and peak to peak values of the measured noises are shown in Table 3.

Table 3: RMS and Peak to Peak measured low frequency noise (0.01Hz to 20Hz)
  RMS [uV] Peak to Peak [mV]
Specification [10uF] 714.0 4.3
PD200_1 565.1 3.7
PD200_2 767.6 3.5
PD200_3 479.9 3.0
PD200_4 615.7 3.5
PD200_5 651.0 2.4
PD200_6 473.2 2.7
PD200_7 423.1 2.3

The Amplitude Spectral Density of the measured noise is now computed and shown in Figure 11.

asd_noise_3uF_warmup.png

Figure 11: Amplitude Spectral Density of the measured noise

5.6 DAC noise measurement

In order not to have any quantization noise, we impose the DAC to output a zero voltage. The gain of the low noise amplifier is adjusted to

The Amplitude Spectral Density \(\Gamma_n(\omega)\) of the measured signal is computed. The Amplitude Spectral Density of \(n_{da}\) can be computed taking into account the gain of the pre-amplifier:

\begin{equation} \Gamma_{n_{da}}(\omega) = \frac{\Gamma_m(\omega)}{|G_a(\omega)|} \end{equation}

And it is verify that the Amplitude Spectral Density of \(n_{da}\) is much larger than the one of \(n_a\):

\begin{equation} \Gamma_{n_{da}} \gg \Gamma_{n_a} \end{equation}

noise_measure_setup_dac.png

Figure 12: Sources of noise in the experimental setup

asd_noise_dac.png

5.7 Total noise measurement

Let’s now analyze the measurement of the setup in Figure 6.

The PSD of the measured noise is computed and the ASD is shown in Figure 14.

win = hanning(ceil(0.5/Ts));

for i = 1:7
    [pxx, f] = pwelch(pd200dac{i}.Vn, win, [], [], Fs);
    pd200dac{i}.f = f;
    pd200dac{i}.pxx = pxx;
end

asd_noise_tot.png

Figure 14: Amplitude Spectral Density of the measured noise and of the individual sources of noise

The output noise of the PD200 amplifier is limited by the noise of the DAC. Having a DAC with lower noise could lower the output noise of the PD200. SSI2V DACs will be used to verify that.

5.8 20bits DAC noise measurement

Let’s now measure the noise of another DAC called the “SSI2V” (doc). It is a 20bits DAC with an output of +/-10.48 V and a very low noise.

The measurement setup is the same as the one in Figure 12.

win = hanning(ceil(0.5/Ts));

[pxx, f] = pwelch(ssi2v.Vn, win, [], [], Fs);
ssi2v.pxx = pxx;
ssi2v.f = f;

The obtained noise of the SSI2V DAC is shown in Figure 15 and compared with the noise of the 16bits DAC. It is shown to be much smaller (~1 order of magnitude).

asd_ssi2v_noise.png

Figure 15: Amplitude Spectral Density of the SSI2V DAC’s noise

Using the SSI2V as the DAC with the PD200 should give much better noise output than using the 16bits DAC. The limiting factor should then be the noise of the PD200 itself.

6 Transfer Function measurement

6.1 Setup

In order to measure the transfer function from the input voltage \(V_{in}\) to the output voltage \(V_{out}\), the test bench shown in Figure 16 is used.

Here are the documentation of the equipment used for this test bench:

For this measurement, the sampling frequency of the Speedgoat ADC should be as high as possible.

setup-dynamics-measurement.png

Figure 16: Schematic of the test bench to estimate the dynamics from voltage input \(V_{in}\) to voltage output \(V_{out}\)

6.2 Maximum Frequency/Voltage to not overload the amplifier

The maximum current is 1A [rms] which corresponds to 0.7A in amplitude of the sin wave.

The impedance of the capacitance is: \[ Z_C(\omega) = \frac{1}{jC\omega} \]

Therefore the relation between the output current amplitude and the output voltage amplitude for sinusoidal waves of frequency \(\omega\): \[ V_{out} = \frac{1}{C\omega} I_{out} \]

Moreover, there is a gain of 20 between the input voltage and the output voltage: \[ 20 V_{in} = \frac{1}{C\omega} I_{out} \]

For a specified voltage input amplitude \(V_{in}\), the maximum frequency at which the output current reaches its maximum value is: \[ \omega_{\text{max}} = \frac{1}{20 C V_{in}} I_{out,\text{max}} \]

\(\omega_max\) as a function of \(V_{in}\) is shown in Figure 17.

max_frequency_voltage.png

Figure 17: Maximum frequency as a function of the excitation voltage amplitude

6.3 Obtained Transfer Functions

Several identifications using sweep sin were performed with input voltage amplitude ranging from 0.1V to 4V.

The obtained frequency response functions are shown in Figure 18. As the input voltage increases, the voltage drop is increasing.

pd200_tf_voltage.png

Figure 18: Transfer function for the PD200 amplitude between \(V_{in}\) and \(V_{out}\) for multiple voltage amplitudes

The small signal transfer function of the amplifier can be approximated by a first order low pass filter.

Gp = 19.95/(1 + s/2/pi/35e3);

The comparison from the model and measurements are shown in Figure 19.

tf_pd200_model.png

Figure 19: Comparison of the model transfer function and the measured frequency response function

7 Conclusion

Table 4: Measured characteristics, Manual characterstics and specified ones
Characteristics Measurement Manual Specification
Input Voltage Range - +/- 10 [V] +/- 10 [V]
Output Voltage Range - -50/150 [V] -20/150 [V]
Gain   20 [V/V] -
Maximum RMS current   0.9 [A] > 50 [mA]
Maximum Pulse current   10 [A] -
Slew Rate   150 [V/us] -
Noise (10uF load)   0.7 [mV RMS] < 2 [mV rms]
Small Signal Bandwidth (10uF load)   7.4 [kHz] > 5 [kHz]
Large Signal Bandwidth (150V, 10uF)   300 [Hz] -

Author: Dehaeze Thomas