ESRF Double Crystal Monochromator - Metrology
-Table of Contents
- --
This report is also available as a pdf.
-- -
-In this document, the metrology system is studied. -First, in Section 1 the goal of the metrology system is stated and the proposed concept is described. -In order to increase the accuracy of the metrology system, two problems are to be dealt with: -
--
-
- The deformation of the metrology frame under the action of gravity (Section 3) -
- The periodic non-linearity of the interferometers (Section 4) -
1. Metrology Concept
--The goal of the metrology system is to measure the distance and default of parallelism orientation between the first and second crystals -
- --Only 3 degrees of freedom are of interest: -
--
-
- \(d_z\) -
- \(r_y\) -
- \(r_x\) -
1.1. Sensor Topology
--In order to measure the relative pose of the two crystals, instead of performing a direct measurement which is complicated, the pose of the two crystals are measured from a metrology frame. -Three interferometers are used to measured the 3dof of interest for each crystals. -Three additional interferometers are used to measured the relative motion of the metrology frame. -
- -Notation | -Meaning | -
---|---|
d |
-“Downstream”: Positive X | -
u |
-“Upstream”: Negative X | -
h |
-“Hall”: Positive Y | -
r |
-“Ring”: Negative Y | -
f |
-“Frame” | -
1 |
-“First Crystals” | -
2 |
-“Second Crystals” | -
-
-Figure 1: Schematic of the Metrology System
-1.2. Crystal’s motion computation
--From the raw interferometric measurements, the pose between the first and second crystals can be computed. -
- --First, Jacobian matrices can be used to convert raw interferometer measurements to axial displacement and orientation of the crystals and metrology frame. -
- --For the 311 crystals: -
- -Notation | -Description | -
---|---|
um |
-Metrology Frame - Upstream | -
dhm |
-Metrology Frame - Downstream Hall | -
drm |
-Metrology Frame - Downstream Ring | -
ur1 |
-First Crystal - Upstream Ring | -
h1 |
-First Crystal - Hall | -
dr1 |
-First Crystal - Downstream Ring | -
ur2 |
-First Crystal - Upstream Ring | -
h2 |
-First Crystal - Hall | -
dr2 |
-First Crystal - Downstream Ring | -
Notation | -Description | -
---|---|
dzm |
-Positive: increase of distance | -
rym |
-- |
rxm |
-- |
dz1 |
-Positive: decrease of distance | -
ry1 |
-- |
rx1 |
-- |
dz2 |
-Positive: increase of distance | -
ry2 |
-- |
rx2 |
-- |
-
-Figure 2: Forward Kinematics for the Metrology frame
--
-Figure 3: Forward Kinematics for the 1st crystal
--
-Figure 4: Forward Kinematics for the 2nd crystal
--Then, the displacement and orientations can be combined as follows: -
-\begin{align} - d_{z} &= + d_{z1} - d_{z2} + d_{zm} \\ - d_{r_y} &= - r_{y1} + r_{y2} - r_{ym} \\ - d_{r_x} &= - r_{x1} + r_{x2} - r_{xm} -\end{align} - --Therefore: -
--
-
- \(d_z\) represents the distance between the two crystals -
- \(d_{r_y}\) represents the rotation of the second crystal w.r.t. the first crystal around \(y\) axis -
- \(d_{r_x}\) represents the rotation of the second crystal w.r.t. the first crystal around \(x\) axis -
-If \(d_{r_y}\) is positive, the second crystal has a positive rotation around \(y\) w.r.t. the first crystal. -Therefore, the second crystal should be actuated such that it is making a negative rotation around \(y\) w.r.t. metrology frame. -
- --The Jacobian matrices are defined as follow: -
-%% Sensor Jacobian matrix for the metrology frame -J_m = [1, 0.102, 0 - 1, -0.088, 0.1275 - 1, -0.088, -0.1275]; - -%% Sensor Jacobian matrix for 1st "111" crystal -J_s_111_1 = [-1, -0.036, -0.015 - -1, 0, 0.015 - -1, 0.036, -0.015]; - -%% Sensor Jacobian matrix for 2nd "111" crystal -J_s_111_2 = [1, 0.07, 0.015 - 1, 0, -0.015 - 1, -0.07, 0.015]; --
-Therefore, the matrix that gives the relative pose of the crystal from the 9 interferometers is: -
-%% Compute the transformation matrix -G_111_t = [-inv(J_s_111_1), inv(J_s_111_2), -inv(J_m)]; - -% Sign convention for the axial motion -G_111_t(1,:) = -G_111_t(1,:); --
- | ur1 [nm] |
-h1 [nm] |
-dr1 [nm] |
-ur2 [nm] |
-h2 [nm] |
-dr1 [nm] |
-um [nm] |
-dhm [nm] |
-drm [nm] |
-
---|---|---|---|---|---|---|---|---|---|
dz [nm] |
--0.25 | --0.5 | --0.25 | --0.25 | --0.5 | --0.25 | -0.463 | -0.268 | -0.268 | -
rx [nrad] |
-13.889 | -0.0 | --13.889 | -7.143 | -0.0 | --7.143 | --5.263 | -2.632 | -2.632 | -
ry [nrad] |
-16.667 | --33.333 | -16.667 | -16.667 | --33.333 | -16.667 | -0.0 | --3.922 | -3.922 | -
-From table 4, we can determine the effect of each interferometer on the estimated relative pose between the crystals.
-For instance, an error on dr1
will have much greater impact on ry
than an error on drm
.
-
2. Relation Between Crystal position and X-ray measured displacement
-2.1. Setup
--
-Figure 5: Schematic of the setup
--Detector: -
- --https://www.baslerweb.com/en/products/cameras/area-scan-cameras/ace/aca1920-40gc/ -
- --Pixel size depends on the magnification used (1x, 6x, 12x). -
- --Pixel size of camera is 5.86 um x 5.86 um. -With typical magnification of 6x, pixel size is ~1.44um x 1.44um -
- --Frame rate is: 42 fps -
-2.2. Relation between second crystal motion and beam motion
-2.2.1. Axial motion of second crystal
--Let’s consider the relation between the \([y, z]\) motion of the beam and the motion of the second crystal \([z^\prime, R_{y^\prime}, R_{x^\prime}]\). -
- - --
-Figure 6: Relation between \(d_{z^\prime}\) motion of the second crystal and vertical motion of the beam
--
-Figure 7: Relation between vertical motion of the second crystal and vertical motion of the output beam
-2.2.2. Ry motion of second crystal
--with \(D_{\text{vlm}} \approx 10\,m\). -
-2.2.3. Rx motion of second crystal
-3. Deformations of the Metrology Frame
--The transformation matrix in Table 4 is valid only if the metrology frames are solid bodies. -
- --The metrology frame itself is experiencing some deformations due to the gravity. -When the bragg axis is scanned, the effect of gravity on the metrology frame is changing and this introduce some measurement errors. -
- --This can be calibrated. -
-3.1. Measurement Setup
--Two beam viewers: -
--
-
- one close to the DCM to measure position of the beam -
- one far away to the DCM to measure orientation of the beam -
-For each Bragg angle, the Fast Jacks are actuated to that the beam is at the center of the beam viewer. -Then, then position of the crystals as measured by the interferometers is recorded. -This position is the wanted position for a given Bragg angle. -
-3.2. Simulations
--The deformations of the metrology frame and therefore the expected interferometric measurements can be computed as a function of the Bragg angle. -This may be done using FE software. -
-3.3. Comparison
-3.4. Test
-aa = importdata("correctInterf-vlm-220201.dat");
-
-figure; -plot(aa.data(:,1), aa.data(:,24)) --
3.5. Measured frame deformation
-data = table2array(readtable('itf_polynom.csv','NumHeaderLines',1)); -th = pi/180*data(:,1); % [rad] -fj = 0.030427 - 10.51e-3./(2*cos(th)); % [m] -rx2 = 1e-9*data(:,2); % [rad] -ry2 = 1e-9*data(:,3); % [rad] -rx1 = 1e-9*data(:,4); % [rad] -ry1 = 1e-9*data(:,5); % [rad] --
figure; -hold on; -plot(180/pi*th, 1e6*detrend(-rx2, 1), '.') -% plot(180/pi*th, detrend(ry2, 1)) -% plot(180/pi*th, detrend(rx1, 1)) -% plot(180/pi*th, detrend(ry1, 1)) -hold off; -xlabel('Bragg Angle [deg]'); -ylabel('Measured $R_x$ [$\mu$rad]') -xlim([10, 75]); --
-
-Figure 8: description
--Strange that there is correlation between Rx and Ry. -
-figure; -hold on; -plot(108/pi*th, 1e9*detrend(rx1, 0), '-', 'DisplayName', '$Rx_1$') -plot(108/pi*th, 1e9*detrend(ry1, 0), '-', 'DisplayName', '$Ry_1$') -hold off; -xlabel('Bragg Angle [deg]'); ylabel('Angle Offset [nrad]'); -legend() --
figure; -hold on; -plot(1e3*fj, detrend(rx2, 1), 'DisplayName', '$Rx_1$') -plot(1e3*fj, detrend(ry2, 1), 'DisplayName', '$Ry_1$') -hold off; -xlabel('Fast Jack Displacement [mm]'); ylabel('Angle Offset [nrad]'); -legend() --
%% Compute best polynomial fit -f_rx2 = fit(180/pi*th, 1e9*rx2, 'poly4'); -f_ry2 = fit(180/pi*th, 1e9*ry2, 'poly4'); -f_rx1 = fit(180/pi*th, 1e9*rx1, 'poly4'); -f_ry1 = fit(180/pi*th, 1e9*ry1, 'poly4'); --
figure; -hold on; -plot(180/pi*th, f_rx2(180/pi*th)) -plot(180/pi*th, f_ry2(180/pi*th)) -plot(180/pi*th, f_rx1(180/pi*th)) -plot(180/pi*th, f_ry1(180/pi*th)) -hold off; --
figure; -hold on; -plot(th, f_rx2(th) - rx2) -plot(th, f_ry2(th) - ry2) -plot(th, f_rx1(th) - rx1) -plot(th, f_ry1(th) - ry1) -hold off; --
figure; -hold on; -plot(th, f(th)) -plot(th, rx2, '.') --
3.6. Test
-filename = "/home/thomas/mnt/data_id21/22Jan/blc13550/id21/test_xtal1_interf/test_xtal1_interf_0001/test_xtal1_interf_0001.h5";
-
-data = struct(); - -data.xtal1_111_u = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM4/data')); -data.xtal1_111_m = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM5/data')); -data.xtal1_111_d = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM6/data')); -data.mframe_u = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM3/data')); -data.mframe_dh = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM4/data')); -data.mframe_dr = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM5/data')); -data.bragg = (pi/180)*double(h5read(filename, '/7.1/instrument/FPGA2_SSIM6/data')); -data.fj_pos = 0.030427 - 10.5e-3./(2*cos(data.bragg)); -data.time = double(h5read(filename, '/7.1/instrument/time/data')); -data.rx = 1e-9*double(h5read(filename, '/7.1/instrument/xtal1_111_rx/data')); -data.ry = 1e-9*double(h5read(filename, '/7.1/instrument/xtal1_111_ry/data')); -data.z = 1e-9*double(h5read(filename, '/7.1/instrument/xtal1_111_z/data')); -data.drx = 1e-9*double(h5read(filename, '/7.1/instrument/xtal_111_drx_filter/data')); -data.dry = 1e-9*double(h5read(filename, '/7.1/instrument/xtal_111_dry_filter/data')); -data.dz = 1e-9*double(h5read(filename, '/7.1/instrument/xtal_111_dz_filter/data')); -data.xtal2_111_u = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM10/data'))+10.5e6./(2*cos(data.bragg)); -data.xtal2_111_m = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM1/data'))+10.5e6./(2*cos(data.bragg)); -data.xtal2_111_d = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM2/data'))+10.5e6./(2*cos(data.bragg)); --
-
-Figure 9: Drifts of the second crystal as a function of Bragg Angle
-figure; -hold on; -plot(108/pi*data.bragg, data.drx) -plot(108/pi*data.bragg, data.dry) -hold off; --
3.7. Repeatability of frame deformation
-filename = "/home/thomas/mnt/data_id21/22Jan/blc13550/id21/test_xtal1_interf/test_xtal1_interf_0001/test_xtal1_interf_0001.h5";
-
-data_1 = struct(); -data_1.time = double(h5read(filename, '/7.1/instrument/time/data')); -data_1.bragg = (pi/180)*double(h5read(filename, '/7.1/instrument/FPGA2_SSIM6/data')); -data_1.fj_pos = 0.030427 - 10.5e-3./(2*cos(data_1.bragg)); -data_1.xtal1_111_u = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM4/data')); -data_1.xtal1_111_m = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM5/data')); -data_1.xtal1_111_d = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM6/data')); -data_1.xtal2_111_u = double(h5read(filename, '/7.1/instrument/FPGA1_SSIM10/data'))+10.5e6./(2*cos(data_1.bragg)); -data_1.xtal2_111_m = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM1/data'))+10.5e6./(2*cos(data_1.bragg)); -data_1.xtal2_111_d = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM2/data'))+10.5e6./(2*cos(data_1.bragg)); -data_1.mframe_u = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM3/data')); -data_1.mframe_dh = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM4/data')); -data_1.mframe_dr = double(h5read(filename, '/7.1/instrument/FPGA2_SSIM5/data')); -data_1.drx = 1e-9*double(h5read(filename, '/7.1/instrument/xtal_111_drx_filter/data')); -data_1.dry = 1e-9*double(h5read(filename, '/7.1/instrument/xtal_111_dry_filter/data')); -data_1.dz = 1e-9*double(h5read(filename, '/7.1/instrument/xtal_111_dz_filter/data')); --
data_2 = struct(); -data_2.time = double(h5read(filename, '/6.1/instrument/time/data')); -data_2.bragg = (pi/180)*double(h5read(filename, '/6.1/instrument/FPGA2_SSIM6/data')); -data_2.fj_pos = 0.030427 - 10.5e-3./(2*cos(data_2.bragg)); -data_2.xtal1_111_u = double(h5read(filename, '/6.1/instrument/FPGA1_SSIM4/data')); -data_2.xtal1_111_m = double(h5read(filename, '/6.1/instrument/FPGA1_SSIM5/data')); -data_2.xtal1_111_d = double(h5read(filename, '/6.1/instrument/FPGA1_SSIM6/data')); -data_2.xtal2_111_u = double(h5read(filename, '/6.1/instrument/FPGA1_SSIM10/data'))+10.5e6./(2*cos(data_2.bragg)); -data_2.xtal2_111_m = double(h5read(filename, '/6.1/instrument/FPGA2_SSIM1/data'))+10.5e6./(2*cos(data_2.bragg)); -data_2.xtal2_111_d = double(h5read(filename, '/6.1/instrument/FPGA2_SSIM2/data'))+10.5e6./(2*cos(data_2.bragg)); -data_2.mframe_u = double(h5read(filename, '/6.1/instrument/FPGA2_SSIM3/data')); -data_2.mframe_dh = double(h5read(filename, '/6.1/instrument/FPGA2_SSIM4/data')); -data_2.mframe_dr = double(h5read(filename, '/6.1/instrument/FPGA2_SSIM5/data')); -data_2.drx = 1e-9*double(h5read(filename, '/6.1/instrument/xtal_111_drx_filter/data')); -data_2.dry = 1e-9*double(h5read(filename, '/6.1/instrument/xtal_111_dry_filter/data')); -data_2.dz = 1e-9*double(h5read(filename, '/6.1/instrument/xtal_111_dz_filter/data')); --
4. Attocube - Periodic Non-Linearity
--Interferometers have some periodic nonlinearity (Thurner et al., 2015). -The period is a fraction of the wavelength (usually \(\lambda/2\)) and can be due to polarization mixing, non perfect alignment of the optical components and unwanted reflected beams (See Ducourtieux, 2018, p. 67 to 69; Thurner et al., 2015). -The amplitude of the nonlinearity can vary from a fraction of a nanometer to tens of nanometers. -
- --In the DCM case, when using Attocube interferometers, the period non-linearity are in the order of several nanometers with a period of \(765\,nm\). -This is inducing some positioning errors which are too high. -
- --In order to overcome this issue, the periodic non-linearity of the interferometers have to be calibrated. -To do so, a displacement is imposed and measured both by the interferometers and by another metrology system which does not have this nonlinearity. -By comparing the two measured displacements, the nonlinearity can be calibration. -This process is performed over several periods in order to characterize the error over the full stroke. -
-4.1. Measurement Setup
--The metrology that will be compared with the interferometers are the strain gauges incorporated in the PI piezoelectric stacks. -
- --It is here supposed that the measured displacement by the strain gauges are converted to the displacement at the interferometer locations. -It is also supposed that we are at a certain Bragg angle, and that the stepper motors are not moving: only the piezoelectric actuators are used. -
- --Note that the strain gauges are measuring the relative displacement of the piezoelectric stacks while the interferometers are measuring the relative motion between the second crystals and the metrology frame. -
- --Only the interferometers measuring the second crystal motion can be calibrated here. -
- --As any deformations of the metrology frame of deformation of the crystal’s support can degrade the quality of the calibration, it is better to perform this calibration without any bragg angle motion. -
- --The setup is schematically with the block diagram in Figure 10. -
- --The signals are: -
--
-
- \(u\): Reference Signal sent to the PI controller (position where we wish to three stacks to be). -The PI controller takes care or controlling to position as measured by the strain gauges such that it is close to the reference position. -
- \(d\): Disturbances affecting the position of the crystals -
- \(y\): Displacement of the crystal as measured by one interferometer -
- \(y_g\): Measurement of the motion in the frame of the interferometer by the strain gauge with some noise \(n_g\) -
- \(y_a\): Measurement of the crystal motion by the interferometer with some noise \(n_a\) -
-
-Figure 10: Block Diagram schematic of the setup used to measure the periodic non-linearity of the Attocube
--The problem is to estimate the periodic non-linearity of the Attocube from the imperfect measurements \(y_a\) and \(y_g\). -
-4.2. Choice of the reference signal
--The main specifications for the reference signal are; -
--
-
- sweep several periods (i.e. several micrometers) -
- stay in the linear region of the strain gauge -
- no excitation of mechanical modes (i.e. the frequency content of the signal should be at low frequency) -
- no phase shift due to limited bandwidth of both the interferometers and the strain gauge -
- the full process should be quite fast -
-The travel range of the piezoelectric stacks is 15 micrometers, the resolution of the strain gauges is 0.3nm and the maximum non-linearity is 0.15%. -If one non-linear period is swept (765nm), the maximum estimation error of the strain gauge is around 1nm. -
- --Based on the above discussion, one suitable excitation signal is a sinusoidal sweep with a frequency of 10Hz. -
-4.3. Repeatability of the non-linearity
--Instead of calibrating the non-linear errors of the interferometers over the full fast jack stroke (25mm), one can only calibrate the errors of one period. -
- --For that, we need to make sure that the errors are repeatable from one period to the other and also the period should be very precisely estimated (i.e. the wavelength of the laser). -
- --Also, the laser wavelength should be very stable (specified at 50ppb). -
- --One way to precisely estimate the laser wavelength is to estimate the non linear errors of the interferometer at an initial position, and then to estimate the non linear errors at a large offset, say 10mm. -
-4.4. Simulation
--Suppose we have a first approximation of the non-linear period. -
-period_est = 765e-9; % Estimated period [m] --
-And suppose the real period of the non-linear errors is a little bit above (by 0.02nm): -
-period_err = 0.02e-9; % Error on the period estimation [m] -period_nl = period_est + period_err; % Period of the non-linear errors [m] --
-The non-linear errors are first estimated at the beginning of the stroke (Figure 11). -
- --
-Figure 11: Estimation of the non-linear errors at the beginning of the stroke
--From this only measurement, it is not possible to estimate with great accuracy the period of the error. -To do so, the same measurement is performed with a stroke of several millimeters (Figure 12). -
- --It can be seen that there is an offset between the estimated and the measured errors. -This is due to a mismatch between the estimated period and the true period of the error. -
- - --
-Figure 12: Estimated non-linear errors at a latter position
--Suppose the non-linear error is characterized by a periodic function \(\mathcal{E}\), to simplify let’s take a sinusoidal function (this can be generalized by taking the fourier transform of the function): -
-\begin{equation} -\mathcal{E}(x) = \sin\left(\frac{x}{\lambda}\right) -\end{equation} --with \(x\) the displacement and \(\lambda\) the period of the error. -
- --The measured error at \(x_0\) is then: -
-\begin{equation} -\mathcal{E}_m(x_0) = \sin\left( \frac{x_0}{\lambda} \right) -\end{equation} --And the estimated one is: -
-\begin{equation} -\mathcal{E}_e(x_0) = \sin \left( \frac{x_0}{\lambda_{\text{est}}} \right) -\end{equation} --with \(\lambda_{\text{est}}\) the estimated error’s period. -
- --From Figure 12, we can see that there is an offset between the two curves. -Let’s call this offset \(\epsilon_x\), we then have: -
-\begin{equation} -\mathcal{E}_m(x_0) = \mathcal{E}_e(x_0 + \epsilon_x) -\end{equation} - --Which gives us: -
-\begin{equation} -\sin\left( \frac{x_0}{\lambda} \right) = \sin \left( \frac{x_0 + \epsilon_x}{\lambda_{\text{est}}} \right) -\end{equation} - --Finally: -
-\begin{equation} -\boxed{\lambda = \lambda_{\text{est}} \frac{x_0}{x_0 + \epsilon_x}} -\end{equation} - --The estimated delay is computed: -
-%% Estimation of the offset between the estimated and measured errors -i_period = stroke > 5e-3-period_nl/2 & stroke < 5e-3+period_nl/2; -epsilon_x = finddelay(nl_errors(i_period), est_errors(i_period)) % [m] --
-Estimated delay x0 is -120 [nm] -- - -
-And the period \(\lambda\) can be estimated: -
-%% Computation of the period [m] -period_fin = period_est * (5e-3)/(5e-3 + d_offset); % Estimated period after measurement [m] --
-The estimated period is 765.020 [nm] -- - -
-And the results confirms that this method is working on paper. -
- --When doing this computation, we suppose that there are at most one half period of offset between the estimated and the measured non-linear (to not have any ambiguity whether the estimated period is too large or too small). -Mathematically this means that the displacement \(x_0\) should be smaller than: -
-\begin{equation} -x_0 < \frac{1}{2} \cdot \lambda \cdot \frac{\lambda}{\epsilon_\lambda} -\end{equation} --With \(\epsilon_\lambda\) the absolute estimation error of the period in meters. -
- --For instance, if we estimate the error on the period to be less than 0.1nm, the maximum displacement is: -
-%% Estimated maximum stroke [m] -max_x0 = 0.5 * 765e-9 * (765e-9)/(0.1e-9); --
-The maximum stroke is 2.9 [mm] --
4.5. Measurements
--We have some constrains on the way the motion is imposed and measured: -
--
-
- We want the frequency content of the imposed motion to be at low frequency in order not to induce vibrations of the structure. -We have to make sure the forces applied by the piezoelectric actuator only moves the crystal and not the fast jack below. -Therefore, we have to move much slower than the first resonance frequency in the system. -
- As both \(y_a\) and \(y_g\) should have rather small noise, we have to filter them with low pass filters. -The cut-off frequency of the low pass filter should be high as compared to the motion (to not induce any distortion) but still reducing sufficiently the noise. -Let’s say we want the noise to be less than 1nm (\(6 \sigma\)). -
-Suppose we have the power spectral density (PSD) of both \(n_a\) and \(n_g\). -
- --
-
[ ]
Take the PSD of the Attocube
-[ ]
Take the PSD of the strain gauge
-[ ]
Using 2nd order low pass filter, estimate the required low pass filter cut-off frequency to have sufficiently low noise
-